6.) Detectors
- six arrays of photoconductive detectors used in SWS:
4 arrays with 12 detectors each in the grating sections
2 arrays with 2 detectors each in the Fabry-Pérot section
- arrays arranged in the direction of dispersion
- they define the SWS band structure (band 1 - 6)
- cover the SWS spectral range 2.38 - 45.2 µm
- detector size determines FOV perpendicular to dispersion
- relative detector size:
d LW ~ 1.5 d SW
and
d FP ~ 2 d SW
- heaters are provided for all detector arrays (FPU at ~ 2 K)
- Si:Sb array is heated to an optimum temperature of 10 K
- Ge:Be array is connected to a cryogenic 2 K heat sink
Table: SWS detector properties
section |
band |
detector |
Temp. |
bias |
noise |
wavelength |
|
|
numb. |
type |
[K] |
[V] |
[µV/s] |
[µm] |
SW |
1 |
12 |
InSb |
3 |
0 |
0.7 |
2.38 - 4.08 |
2 |
12 |
Si:Ga |
3 |
12 |
1.5 |
4.08 - 12.0 |
LW |
3 |
12 |
Si:As |
4.5 |
3 |
3.0 |
12.0 - 29.0 |
4 |
12 |
Ge:Be |
2 |
0.6 |
2.0 |
29.0 - 45.2 |
FP |
5 |
2 |
Si:Sb |
10 |
1.1 |
1.1 |
11.4 - 26.0 |
6 |
2 |
Ge:Be |
2 |
0.6 |
3.0 |
26.0 - 44.5 |
- cold read-out electronics:
integrating pre-amplifiers with non-destructive read-out,
employing heated JFET's
read-out at a rate of 24 Hz
signal of all 52 detectors
part of Focal Plane Unit (FPU)
- warm amplifier electronics:
52 identical main amplifiers
multiplexer and ADC circuitry
outside cryo-harness
- 3 reset intervals: tr = 1 and 2 sec,
also 4 sec for band 1
determined from observer's input
longer reset intervals for faint sources
reset pulse discharges capacitor (basically short-circuited)
- 3 gain settings: 1x, 4x or 16x
determined from observer's input
no saturation, if source flux density < 50,000 - 100,000 Jy
and flux density estimate accurate to a factor of four