Thermal Vacuum Chamber II

This thermal vacuum Chamber significantly helps the outgassing of assemblies, cables, plastic elements, etc. This should prevent cleaner environments from contaminated with condensable materials during operation or further testing.

Overall:

Maximum load (Ø x H):

800 x 1200 (in mm)

Realisation of  electrical process:

Realisation on request

Thermaldaten:

Minimum temperature:

Room temperature (no cooling)

Maximal temperature:

+ 120 ° C

Temperature control:

Manual mode or  programme mode

Dynamic temperature control:

0 - 120 K/h

Temperature monitoring:

Hybrid recorder with 10 freely available
temperature measuring channels and 9 freely
available voltage measurement channels

Vacuum data:

Backing pumps:

Scroll Vacuum Pumps

High vacuum pump:

Turbo molecular pump

Final pressure (empty recipient)

6.0 E - 7 mbar at 20 ° C after 2 days of
pumping time

Partial pressure control:

Mass spectrometer Balzers QMG 064 to 63
mass amu