Thermal Vacuum Chamber II
This thermal vacuum Chamber significantly helps the outgassing of assemblies, cables, plastic elements, etc. This should prevent cleaner environments from contaminated with condensable materials during operation or further testing.
Overall: |
|
Maximum load (Ø x H): |
800 x 1200 (in mm) |
Realisation of electrical process: |
Realisation on request |
Thermaldaten: |
|
Minimum temperature: |
Room temperature (no cooling) |
Maximal temperature: |
+ 120 ° C |
Temperature control: |
Manual mode or programme mode |
Dynamic temperature control: |
0 - 120 K/h |
Temperature monitoring: |
Hybrid recorder with 10 freely available |
Vacuum data: |
|
Backing pumps: |
Scroll Vacuum Pumps |
High vacuum pump: |
Turbo molecular pump |
Final pressure (empty recipient) |
6.0 E - 7 mbar at 20 ° C after 2 days of |
Partial pressure control: |
Mass spectrometer Balzers QMG 064 to 63 mass amu |